Calibration of a superconducting gravimeter with an absolute atom gravimeter

verfasst von
S. Merlet, Pierre Gillot, Bing Cheng, Romain Karcher, Almazbek Imanaliev, Ludger Timmen, Franck Pereira dos Santos
Abstract

We present a 27-day-long common view measurement of an absolute cold atom gravimeter (CAG) and a relative iGrav superconducting gravimeter, which we use to calibrate the iGrav scale factor. This allowed us to push the CAG long-term stability down to the level of 0.5 nm s- 2. We investigate the impact of the duration of the measurement on the uncertainty in the determination of the correlation factor and show that it is limited to about 3‰ by the coloured noise of our cold atom gravimeter. A 3-day-long measurement session with an additional FG5X absolute gravimeter allows us to directly compare the calibration results obtained with two different absolute meters. Based on our analysis, we expect that with an improvement of its long-term stability, the CAG will allow to calibrate the iGrav scale factor to better than the per mille level (1σ level of confidence) after only 1 day of concurrent measurements for maximum tidal amplitudes.

Organisationseinheit(en)
Institut für Erdmessung
SFB 1464: Relativistische und quanten-basierte Geodäsie (TerraQ)
Externe Organisation(en)
SYRTE - Observatoire de Paris
Typ
Artikel
Journal
Journal of geodesy
Band
95
ISSN
0949-7714
Publikationsdatum
10.05.2021
Publikationsstatus
Veröffentlicht
Peer-reviewed
Ja
ASJC Scopus Sachgebiete
Geophysik, Geochemie und Petrologie, Computer in den Geowissenschaften
Elektronische Version(en)
https://doi.org/10.1007/s00190-021-01516-6 (Zugang: Geschlossen)
 

Details im Forschungsportal „Research@Leibniz University“