Comparison between two mobile absolute gravimeters
Optical versus atomic interferometers
- authored by
- S. Merlet, Q. Bodart, N. Malossi, A. Landragin, F. Pereira Dos Santos, O. Gitlein, L. Timmen
- Abstract
We report a comparison between two absolute gravimeters: the LNE-SYRTE cold atom gravimeter and FG5#220 of Leibniz Universität of Hannover. They rely on different principles of operation: atomic and optical interferometry. Both are movable which enabled them to participate in the last International Comparison of Absolute Gravimeters (ICAG’09) at BIPM. Immediately after, their bilateral comparison took place in the LNE watt balance laboratory and showed an agreement of (4.3 ± 6.4)µGal.
- Organisation(s)
-
Institute of Geodesy
- External Organisation(s)
-
Centre national de la recherche scientifique (CNRS)
University of Pisa
- Type
- Article
- Journal
- METROLOGIA
- Volume
- 47
- Pages
- L9-L11
- ISSN
- 0026-1394
- Publication date
- 18.06.2010
- Publication status
- Published
- Peer reviewed
- Yes
- ASJC Scopus subject areas
- Engineering(all)
- Electronic version(s)
-
https://doi.org/10.48550/arXiv.1005.0357 (Access:
Open)
https://doi.org/10.1088/0026-1394/47/4/L01 (Access: Closed)
-
Details in the research portal "Research@Leibniz University"